10. Operation Messages

E114

Invalid Measurement !

Make another measurement. If further measurement proved to be futile, check staff position and Inverse 

Staff setting, check the lighting condition at the staff and stray light, check focusing and targeting, check 

if sufficient length of barcode in the field of view.

E115

Temperature Sensor 

Error !

Cover the objective telescope with a hand and switch ON the instrument. Hardware communication failed.

E116

Adjustment Error !

Carry out the adjustment with guided steps, make sure instrument is level and staff is truly vertical in 

normal position. Collimation is out of correction range.

E117

BM Change Not Allowed!

Exit to default measurement mode by pressing HEIGHT/DISTANCE key, and change the Benchmark 

elevation in the INPUT BM menu mode.

E119

Staff Blocked

Not enough barcode length for measurement.

E120

Image sensor Error!

Contact services.

E121

Adjustment Inverse Staff 

Not Allowed!

Check the staff orientation and staff setting.

E123

PtID Change Not Allowed Exit the message by pressing ESC key.

No.

Error message

Counter measure / causes

Operation message

Counter measure / remark

Start Tracking!

Tracking mode starts.

Stop Tracking!

Tracking mode stops.

Tracking Hold!

Press measure key for 3 seconds to restart tracking mode. Tracking will hold after 10 unsuccessful meas-

urements.

Abort Measurement!

Current measuring process is terminated.

Downloading Data! 

Downloading of data from the internal memory to an external device is in progress.