18

Error Messages

Sprinter 150/150M/250M - 1.0.0en

EN
DE
FR
ES

NO

SV

FI

DA

IT

PT
NL

9.

Error Messages

No.

Error message

Counter measure / causes

E99

System Error, Contact 

Services !

Hardware faults or file errors or adjustment errors or setting errors rendering instrument not working prop-

erly. 

E100

Low Battery !

Change to new or freshly charged batteries. 

E101

Point No. Not Incre-

mented!

Change the PtID. Max. PtID is 99999999 and do not end any 8 characters string with an alphabet char-

acter.

E102

Too Bright !

Darken staff or reduce lighting on staff or shade the objective telescope.

E103

Too Dark !

Light up staff uniformly.

E104

No Staff !

Check targeting.

E105

Invalid Entry! 

Check the entry / input.

E106

Out Of Level !

Level the instrument. 

E107

Memory Full!

Set internal memory to OFF and continue measurement without recording OR download stored data to 

an external device and continue measurement with internal memory ON after deleting all recorded data 

in the internal memory.

E108

Data File Error !

Data file error.

E109

Low Memory!

Prepare to download data to external device, in order to continue further measurement with recording ON 

after deleting all recorded data in the internal memory.

E110

Target Too Close !

Move staff or instrument further apart.

E111

Target Too far !

Move staff or instrument closer together.

E112

Too Cold !

Stop working, external temperature is outside the instrument operating temperature. 

E113

Too Hot !

Stop working, external temperature is outside the instrument operating temperature. 