8.

Check & Adjust

8.1 Electronic Collimation Adjustment

To activate "Adjustment" program, go to Menu\Adjustment.

Step 1:

 Aim at Staff A and press MEAS key. Measurement display, 

press ENTER key to accept.

Step 2:

 Aim at Staff B and press MEAS key. Measurement display, 

press ENTER key to accept.

Now shift the Sprinter towards staff A and set it up at about 3 m to 

staff A.

Step 3:

 Aim at Staff B and press MEAS key. Measurement display, 

press ENTER key to accept.

Step 4:

 Aim at Staff A and press MEAS key. Measurement display, 

press ENTER key to accept.

The new electronic collimation error is displayed. To accept new 

correction, press ENTER key, otherwise press ESC key to decline 

adjustment result.

)

Optical collimation error can be corrected by adjusting the 

reticle.

B

A ~15m

~15m

B

A

~3m

~27m