17

Check & Adjust

Sprinter 150/150M/250M - 1.0.0en

EN
DE
FR
ES

NO

SV

FI

DA

IT

PT
NL

8.2 Circular Level

8.3 Optical Collimation / Reticle Ad-

justment

If the collimation error exceeds 3 mm over 60 m distance, the colli-

mation needs to be adjusted.

Step

Description

1.

Level instrument. 

2.

Turn instrument by 180°.

3.

Center bubble if it extends beyond the centering circle.

4.

Correct half of the error with the Allen key.

)

Repeat steps 1 to 4 until the circular level bubble is 

centered at any random telescope pointing. 

r_18

1.

2.

3.

4.

Step

Description

1.

Turn Allen key until design value is reached.

2.

Check collimation.